Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users

  • Christine Morken*
  • , Joakim Hekland
  • , Johan Olausson
  • , Johan Baden Birk-Korch
  • , Heidi Berghäll
  • , Johannes Østrem Fjøse
  • , Gunhild Øygard Fosse
  • , Sander Johannes Thorbjørnsen Guttorm
  • , Gustav Hallén
  • , Anne Haapala
  • , Tine Lise Kalleklev
  • , Stiina Kamula
  • , Dår Kristian Kur
  • , Kristiina Kurg
  • , Cindhya Sithiravel
  • , Anne Elisabeth Solsvik
  • , Ricardas Stonys
  • , Kristel Virtanen
  • , Elisabeth Wiik Vigerust
  • , Anne Stavelin
  • *Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

1 Citation (Scopus)

Abstract

Diabetes is a growing global health concern affecting millions worldwide. Glucose measurement remains a cornerstone in the diagnosis, monitoring, and management of this disease. Considering that numerous point-of-care (POC) devices for glucose measurements are available, systematic evaluation of their analytical performance and user-friendliness is essential. This study presents the findings of an independent evaluation of the cobas pulse glucose meter (Roche Diagnostics GmbH) conducted by the Scandinavian evaluation of laboratory equipment for point of care testing (SKUP). The study, which assessed both the analytical performance and user-friendliness, included 217 participants with diabetes and was conducted in both a hospital laboratory and primary healthcare settings. Capillary whole blood measurements were compared to a plasma glucose hexokinase method at Vejle hospital, Denmark. Certified reference materials from the National Institute of Standards and Technology (NIST) were used to correct all sample results. Repeatability and accuracy were assessed against predefined analytical performance specifications, and user-friendliness was evaluated using a structured questionnaire developed by SKUP. The cobas pulse demonstrated good repeatability (CV: 1.8–4.0%) and fulfilled the specification for accuracy, with minimal bias (maximum +0.16 mmol/L in the 7–10 mmol/L range). Haematocrit had no significant effect within the tested range (5–70%). User-friendliness was rated satisfactory, attributed to intuitive operation, rapid analysis times, and effective quality control features. In conclusion, the cobas pulse glucose meter met the performance criteria for professional use in clinical and primary care settings, supporting its suitability in near-patient glucose monitoring.

Original languageEnglish
JournalScandinavian Journal of Clinical and Laboratory Investigation
Volume85
Issue number6
DOIs
Publication statusPublished - 2025
Publication typeA1 Journal article-refereed

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Blood glucose
  • point-of-care testing
  • primary health care
  • quality control
  • surveys and questionnaires
  • validation study

Publication forum classification

  • Publication forum level 1

ASJC Scopus subject areas

  • Clinical Biochemistry

Fingerprint

Dive into the research topics of 'Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users'. Together they form a unique fingerprint.

Cite this