The effect of inhalation delay on the aerodynamic particle size distribution of fluticasone propionate and ciclesonide using pMDI and valved holding chambers

Research output: Contribution to journalArticleScientificpeer-review

5 Downloads (Pure)

Abstract

Delayed inhalation when using pressurised metered-dose inhalers (pMDIs) for inhaled corticosteroids (ICS) is a common technique error. Valved holding chambers (VHCs) can mitigate its impact, but the effect of delays on drug delivery from suspension versus solution formulations remains poorly understood. We compared the doses of fluticasone propionate (FP) and ciclesonide (CIC) delivered to an anatomical adult throat model and a Next Generation Impactor as particles 1–5 μm and under 1 μm in diameter, using AeroChamber (AC), EasyChamber (EC), and OptiChamber Diamond (OD) VHCs with inhalation delays of 0, 1, 3, and 5 s. A breathing simulator was used to produce a single, adult-type inhalation. Throat deposition gradually decreased with longer inhalation for both FP and CIC, from an average of 3.3 %–2.2 % and 1.0 %–0.5 % of the label claim, respectively. Similarly, deposition of 1–5 μm particles declined from 30 % to 28 % for FP and from 33 % to 25 % for CIC. In contrast, deposition of particles smaller than 1 μm were relatively unaffected by inhalation delays. In conclusion, increasing the inhalation delay up to 5 s slightly reduced the respirable deposition of FP and CIC particles when using a VHC, but these reductions are unlikely to be clinically meaningful due to small differences in absolute dose. The smallest particles likely remained unaffected by the delay due to their low tendency to settle. Differences in performance between AC, EC, and OD during the delay were likely too minor to influence treatment outcomes.

Original languageEnglish
Article number108258
JournalRespiratory Medicine
Volume247
DOIs
Publication statusPublished - Oct 2025
Publication typeA1 Journal article-refereed

Keywords

  • Asthma
  • Drug delivery
  • Fine particle dose
  • Inhaled corticosteroid
  • Spacer
  • Valved holding chamber

Publication forum classification

  • Publication forum level 1

ASJC Scopus subject areas

  • Pulmonary and Respiratory Medicine

Fingerprint

Dive into the research topics of 'The effect of inhalation delay on the aerodynamic particle size distribution of fluticasone propionate and ciclesonide using pMDI and valved holding chambers'. Together they form a unique fingerprint.

Cite this