Kuvaus
Experiments were done to determine the pressure-strain dependency of a silicone membrane in a custom made stretching device. 1 µm Ø Dragon Green fluorescent polystyrene microbeads (Bangs Laboratories, Inc., Fishers, IN, USA) were adsorbed to the surface of the silicone membrane. ZEISS Axio Scope microscope with A-Plan 10x/0.25 Ph 1 objective was used to take image sequences of membrane deformation. The membrane was deformed by decreasing the pressure from -4 mbar to -350 mbar 75 mbar steps. The microscope was manually refocused after each step. Within one strain measurement, three locations near the edge of the cell culture well (imaging locations locations 1-3) and one location in the center (imaging location4) were measured. Three to five repeated image sequences were taken at each location.For analysis, Particle Tracker 2D/3D plugin (Mosaic) [29] in Fiji [30] was used to track the trajectories of the beads (data not shown).
| Koska saatavilla | 7 toukok. 2022 |
|---|---|
| Julkaisija | ScienceDB |
Field of science, Statistics Finland
- 318 Lääketieteen bioteknologia
- 1182 Biokemia, solu- ja molekyylibiologia
Tutkimustuotos
- 1 Artikkeli
-
Pneumatic equiaxial compression device for mechanical manipulation of epithelial cell packing and physiology
Peussa, H., Kreutzer, J., Mäntylä, E., Mäki, A., Nymark, S., Kallio, P. & Ihalainen, T., 3 kesäk. 2022, julkaisussa: PLoS ONE. 17, 6, e0268570.Tutkimustuotos: Artikkeli › Tieteellinen › vertaisarvioitu
Open accessTiedosto9 Sitaatiot (Scopus)41 Lataukset (Pure)
Tietoaineistot
-
FWHM microscopy data
Peussa, H. (Creator), ScienceDB, 17 toukok. 2022
DOI - pysyväislinkki: 10.57760/sciencedb.01764
Tietoaineisto: Dataset
-
Stretch tests -4 mbar to -600 mbar
Peussa, H. (Creator), ScienceDB, 7 toukok. 2022
DOI - pysyväislinkki: 10.57760/sciencedb.01743
Tietoaineisto: Dataset
Siteeraa tätä
- DataSetCite