T1 - Comparison of Continuous Non-Invasive Blood Pressure Monitors Finapres Nova and Caretaker 4 During Rest and Laboratory Interventions
AU - Sytelä, Siri
AU - Haapatikka, Mira
AU - Vehkaoja, Antti
PY - 2023
Y1 - 2023
N2 - Aim: There is a significant demand for a reliable continuous non-invasive blood pressure monitor which could be used to detect fatal changes in blood pressure for example during and after surgical operations. Thus, this study aimed to find out if two continuous non-invasive blood pressure monitors, Finapres Nova and Caretaker 4, react similarly to fast changes in blood pressure. Methods: 21 healthy young adult subjects were recruited for measurements which included four rest phases and three laboratory intervention tests intended to induce fast changes in blood pressure. The tests were hand grip test, cold pressor test, and Valsalva maneuver. Results: Mean discrepancies and standard deviations for systolic and diastolic blood pressures measured with Finapres and Caretaker were -1.6 ± 8.4 mmHg and -2.3 ± 5.3 mmHg during rest. Caretaker reacted to the blood pressure changes caused by the interventions significantly less than Finapres (p=0. 004 for systolic and p=0. 006 for diastolic blood pressure during cold test and p
AB - Aim: There is a significant demand for a reliable continuous non-invasive blood pressure monitor which could be used to detect fatal changes in blood pressure for example during and after surgical operations. Thus, this study aimed to find out if two continuous non-invasive blood pressure monitors, Finapres Nova and Caretaker 4, react similarly to fast changes in blood pressure. Methods: 21 healthy young adult subjects were recruited for measurements which included four rest phases and three laboratory intervention tests intended to induce fast changes in blood pressure. The tests were hand grip test, cold pressor test, and Valsalva maneuver. Results: Mean discrepancies and standard deviations for systolic and diastolic blood pressures measured with Finapres and Caretaker were -1.6 ± 8.4 mmHg and -2.3 ± 5.3 mmHg during rest. Caretaker reacted to the blood pressure changes caused by the interventions significantly less than Finapres (p=0. 004 for systolic and p=0. 006 for diastolic blood pressure during cold test and p
U2 - 10.22489/CinC.2023.171
DO - 10.22489/CinC.2023.171
M3 - Conference contribution
T3 - Computing in Cardiology
BT - 2023 Computing in Cardiology (CinC)
PB - Computing in Cardiology
T2 - Computing in Cardiology
Y2 - 1 October 2023 through 4 October 2023
ER -