Cone beam CT doses in radiotherapy patient positioning in Finland: prostate treatments

Teemu Siiskonen, Saara Alenius, Tiina Seppälä, Joonas Tikkanen, Miia Nadhum, Jarkko Ojala

Tutkimustuotos: ArtikkeliScientificvertaisarvioitu

Abstrakti

Imaging parameters, frequencies and resulting patient organ doses in treatments of prostate cancer were assessed in Finnish radiotherapy centres. Based on a questionnaire to the clinics, Monte Carlo method was used to estimate organ doses in International Commission on Radiological Protection standard phantom for prostate, bladder, rectum and femoral head. The results show that doses from cone beam computed tomography imaging have reduced compared to earlier studies and are between 3.6 and 34.5 mGy per image for the above-mentioned organs and for normal sized patients. There still is room for further optimization of the patient exposure, as many centres use the default imaging parameters, and the length of the imaged region may not be optimal for the purpose.