Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer

Annele Heikkilä, Eeva Boman, Maija Rossi, Antti Vanhanen, Mikko Mankinen, Michiel Postema, Tuomas Koivumäki

Tutkimustuotos: ArtikkeliScientificvertaisarvioitu

3 Lataukset (Pure)

Abstrakti

Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of 1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
AlkuperäiskieliEnglanti
Artikkeli103203
Sivumäärä7
JulkaisuPhysica Medica
Vuosikerta117
DOI - pysyväislinkit
TilaJulkaistu - 2024
OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä

Julkaisufoorumi-taso

  • Jufo-taso 1

!!ASJC Scopus subject areas

  • Oncology
  • Radiation

Sormenjälki

Sukella tutkimusaiheisiin 'Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Siteeraa tätä