Effect of Operation Pressure on the Response of ePNC Particle Number Concentration Sensor

Elmeri Laakkonen, Anssi Arffman, Antti Rostedt, Jorma Keskinen

Tutkimustuotos: ArtikkeliTieteellinenvertaisarvioitu

2 Sitaatiot (Scopus)
8 Lataukset (Pure)

Abstrakti

In this article, we present a method that aims toward an ideal number concentration response of a diffusion charger and diffusion collector-based particle sensor (Dekati ePNC) by using reduced pressure conditions in the sensor. A model is derived for the charger and diffusion collector stage of the sensor, as well as for the total particle number concentration (PNC) response. The derived model is validated with experimental characterization measurements in different operation pressures. The experimental results show that reduced operation pressure makes the sensor response less dependent on particle size, approaching pure particle number response. The obtained results improve the feasibility of the diffusion charging-based method in the measurement of PNC.

AlkuperäiskieliEnglanti
Sivut798-805
Sivumäärä8
JulkaisuIEEE Sensors Journal
Vuosikerta24
Numero1
DOI - pysyväislinkit
TilaE-pub ahead of print - 20 marrask. 2023
OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä

Julkaisufoorumi-taso

  • Jufo-taso 2

!!ASJC Scopus subject areas

  • Instrumentation
  • Electrical and Electronic Engineering

Sormenjälki

Sukella tutkimusaiheisiin 'Effect of Operation Pressure on the Response of ePNC Particle Number Concentration Sensor'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Siteeraa tätä