Sensitive multi-species photoacoustic gas detection based on mid-infrared supercontinuum source and miniature multipass cell

Tommi Mikkonen, Tuomas Hieta, Goery Genty, Juha Toivonen

Tutkimustuotos: ArtikkeliTieteellinenvertaisarvioitu

12 Sitaatiot (Scopus)
10 Lataukset (Pure)

Abstrakti

We report multipass broadband photoacoustic spectroscopy of trace gases in the mid-infrared. The measurement principle of the sensor relies on supercontinuum-based Fourier transform photoacoustic spectroscopy (FT-PAS), in which a scanning interferometer modulates the intensity of a mid-infrared supercontinuum light source and a cantilever microphone is employed for sensitive photoacoustic detection. With a custom-built external Herriott cell, the supercontinuum beam propagates ten times through a miniature and acoustically non-resonant gas cell. The performance of the FT-PAS system is demonstrated by measuring the fundamental C–H stretch bands of various hydrocarbons. A noise equivalent detection limit of 11 ppb is obtained for methane (40 s averaging time, 15 μW/cm−1 incident power spectral density, 4 cm−1 resolution), which is an improvement by a factor of 12 compared to the best previous FT-PAS systems. High linearity and good stability of the sensor provide reliable identification of individual species from a gas mixture with strong spectral overlap, laying the foundation for sensitive and selective multi-species detection in small sample volumes.
AlkuperäiskieliEnglanti
Sivut19481-19487
JulkaisuPhysical Chemistry Chemical Physics
Numero32
Varhainen verkossa julkaisun päivämäärä1 elok. 2022
DOI - pysyväislinkit
TilaJulkaistu - 2022
OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä

Julkaisufoorumi-taso

  • Jufo-taso 2

Sormenjälki

Sukella tutkimusaiheisiin 'Sensitive multi-species photoacoustic gas detection based on mid-infrared supercontinuum source and miniature multipass cell'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Siteeraa tätä