The electrical impedance measurements of dry electrode materials for the ECG measuring after repeated washing

    Tutkimustuotos: ArtikkeliScientificvertaisarvioitu

    19 Sitaatiot (Scopus)

    Abstrakti

    Purpose

    The purpose of this research was to offer more reliable dry electrode materials for long-term measuring and determine how repeated machine washing affects the measured impedance and surface resistance of the sample electrodes. The aim was to manufacture electrodes that could be used for the measurement of ECG. Skin friendly, metal sheet type, electrodes could be a solution.

    Design/methodology/approach

    In addition to two conventional electrodes already used in heart rate belts, the authors prepared and tested three different sheet metal electrodes. Three 20 mm diameter electrodes were manufactured from the following materials: silvered knit, conductive polymer, stainless steel, silver and platinum. Electrode Impedance was measured at seven frequencies from 1 Hz to 1 MHz, by placing two electrodes face-to-face. Measurements were done on unused electrodes and after multiple machine washes at 40°C.

    Findings

    Analysis of the measurements indicates that with every material tested the impedances are elevated after repeated washes. All metallic materials have impedances in the range of 0.01 to 4.5Ω. Metal sheet electrodes can be integrated comfortably into the textile and they endure textile maintenance without loss of electrical properties.

    Practical implications

    Metal sheet electrodes function well in long-term vital signs monitoring, they provide a reliable signal and they are resistant to maintenance. For the reasons described in this research they can be used as a long-term wearable sensor.

    Originality/value

    Novel electrode material for long-term measuring research is important in many disciplines such as healthcare and apparel manufacturing. These findings suggest that pure metal electrodes are better than conductive textiles in long-term measuring.
    AlkuperäiskieliEnglanti
    Sivut59-71
    JulkaisuResearch Journal of Textile and Apparel
    Vuosikerta21
    Numero1
    DOI - pysyväislinkit
    TilaJulkaistu - helmik. 2017
    OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä

    Julkaisufoorumi-taso

    • Jufo-taso 1

    Sormenjälki

    Sukella tutkimusaiheisiin 'The electrical impedance measurements of dry electrode materials for the ECG measuring after repeated washing'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

    Siteeraa tätä