Abstrakti
We study the wireless powering of intra-abdominal implants using a wearable around-the-body loop antenna. The study includes the wireless link characterization and modelling of the specific absorption rate (SAR) obtained by an anatomical human body model and a simplified homogenous elliptic cylinder model. We also assess the robustness of the system in terms of the power gain uniformity at the cross-sectional plane of the abdominal cavity. Overall, our system achieves the link power efficiency of −6.6 dB at 7.25 MHz to an implant (2.3-by-2.3 cm square loop) at the center of the abdominal cavity. According to SAR analysis, under the 1.6 W/kg limit, we can transmit 530 mW and the maximum power available at the implant is 149 mW with the voltage amplitude of 3.6 V. Moreover, the voltage at the implant remains above 1 V for transmission power levels down to 10% of the maximum.
Alkuperäiskieli | Englanti |
---|---|
Otsikko | 2019 International Conference on Electromagnetics in Advanced Applications (ICEAA) |
Kustantaja | IEEE |
Sivut | 0055-0059 |
Sivumäärä | 5 |
ISBN (elektroninen) | 978-1-7281-0563-5 |
ISBN (painettu) | 978-1-7281-0564-2 |
DOI - pysyväislinkit | |
Tila | Julkaistu - 24 lokak. 2019 |
OKM-julkaisutyyppi | A4 Artikkeli konferenssijulkaisussa |
Tapahtuma | International Conference on Electromagnetics in Advanced Applications - Kesto: 1 tammik. 1900 → … |
Conference
Conference | International Conference on Electromagnetics in Advanced Applications |
---|---|
Ajanjakso | 1/01/00 → … |
Julkaisufoorumi-taso
- Jufo-taso 1